Product

AvCount Lube Low-Visc Particle Counter - SA1910-0

Methods
ASTM D6876
ASTM D7619
ASTM D7647
IP 565
GJB 420-1987/420-A-1996/420B-2006
IEC 60970
ISO 4406
ISO 60970

The AvCount Lube Low-Visc particle counter determines particle concentration in oils with a viscosity up to 64 mm2/s. FFKM seals have been fitted to resist aggressive oils.

• Simple operation
• Ideal for handling lubricating and hydraulic oils
• 3 minute test time
• Programmable via PC

With a simple test procedure and user friendly features, operator skill or expertise is kept to a minimum.

AvCount Lube Low Visc can hold up to 3 test methods with custom methods created on a PC before being uploaded using the supplied ProTrend control and analysis software.

Test progress and results for up to 6 size bands are displayed as cumulative particles/ml and ISO 4406, NAS 1638, AS4059 cleanliness codes.

Supplied with: sample probe, waste pipe, universal voltage power supplies, USB cable, calibration certificate and USB memory stick containing handbooks and ProTrend software for test method programming and data download.

AvCount Lube Low-Visc Particle Counter - SA1910-0

Features

  • Calibrated to ISO 11171, traceable to NIST SRM 2806d
  • Supplied with ASTM D7647, ISO 60970 and ISO 4406 loaded as standard
  • FFKM seals compatible with synthetic oils
  • Suitable for fluids up to 64 mm2/s viscosity
  • Precise and reliable results
  • Ultra compact rugged, steel case suitable for portable use
  • Data storage for up to 600 measurements
  • Internal Double Pump System
  • USB connection for PC connectivity

Information

Details

Calibration & Verification
AvCount Lite utilises 16 point calibration. Calibration can be carried out either using NIST traceable standard solutions or by interfacing with a primary calibrated ‘Master’ AvCount Lite. Calibration can be carried out at a user site or at our authorised regional laboratories. Seta authorised training is required to perform a calibration. Verification can be performed using AvCount Verification Material.

ProTrend (SA1810-0) is supplied with the instrument and can report results to ISO 4406, NAS 1638 and SAE AS4059 (tables 1 and 2) cleanliness codes. Using the software, you can program other test methods and cleanliness codes with up to 16 size bands.

Details

IP 565 Determination of the level of cleanliness of aviation turbine fuels – portable automatic particle counter method
ASTM D7619 Method for counting and sizing particles in light and middle distillate fuels, and liquid bio fuels, by automatic particle counter
GJB 420-1987/420-A-1996/420B-2006 Fluids contamination testing
ASTM D7647 Standard test method for automatic particle counting of lubricating and hydraulic fluids using dilution techniques to eliminate the contribution of water and interfering soft particles by lght extinction
ISO 60970 Insulating liquids - Methods for counting and sizing particles
ISO 4406 Hydraulic fluid power - Fluids - Method for coding the level of contamination by solid particles
IEC 60970 Insulating Liquids, Methods for Counting and Sizing Particles
ASTM D6876 Standard Test Method for Particle Count in Mineral Insulating Oil using Automatic Optical Particle Counters

Details

Particle Size Range4 – 100 µm(c) (in accordance with ISO 11171)
Number of Size BandsUp to 16 when connected to Protrend, 6 when used as a stand-alone
Maximum Particle Count Per ml60 000 particles/ml
Sample Viscosity (Max)64 mm2/s
Sample Temperature Range0 – 70 °C
Results600 internal, unlimited on PC
Sample Flow Rate30 ml/min ±5 ml/min
Total Sample Volume Used80 ml (includes rinse cycles) for ASTM D7647, 20 ml other methods
Power Supply100 - 240 Vac, 50/60 Hz
Size (HxWxD)27 mm × 32 mm × 28 mm
Weight7 kg
CCCN CodeTariff 90278990