Product
Seta Auto-Oxidation Stability Bath - 15400-5
This unit measures the induction period for oxygen in gasoline, and allows the weight of gum and residue in aviation fuel to be determined by direct weighing.
Comprising a heated water bath with reflux condenser, a cold water plunge tank for cooling test vessels after each test, and an integral system for charging oxidation vessels prior to test. The Seta Auto-Oxi Control System (15452-2) provides automatic monitoring of up to four pressure test vessels. High quality pressure transducers and system components ensure reliable operation and fully traceable results
Seta Auto-Oxidation Stability Bath - 15400-5
Features
- Operating temperature 100°C
- Water Bath with integral condenser
- 4 x test stations
- Cold plunge bath
- Seta Auto-Oxi Automatic Pressure Monitoring
- Integral vessel charge system
- Constant level device
Information
Details
The Seta Auto-Oxi Control System (15452-2) is programmed with pressure / time test profiles for ASTM D525, D873, IP 40 and ISO 7536 allowing rapid test selection and start. Alternatively custom test profiles are easily programmed by the user via the ‘Settings’ tab on the display screen.
Sample / operator details can also be input.
The recorded pressure status of each test vessel being monitored is displayed and the software automatically follows the pressure check and test sequence laid down in the test method, the displaying shows the ongoing test status of each vessel as well as the duration of the test remaining.Once a test starts, the system is designed for unattended operation and automatically reports the induction period (pressure drop) of each sample under test.
Results can be printed or downloaded via a USB interface.
Supplied with:
Auto-Oxi PC software
USB Data Cable
USB Thermometer
Oxygen charging system
Mains lead
Instruction manual
Note: IBM compatible PC required with 2 USB ports.
Details
ASTM D873 | Standard test method for oxidation stability of aviation fuels (potential residue method) |
---|---|
ASTM D5304 | Oxidation Stability of Gasoline (Induction Period Method) |
IP 40 | Determination of oxidation stability of gasoline - Induction period method |
ASTM D525 | Standard test method for oxidation stability of gasoline (induction period method) |
BS EN ISO 7536 & BS 2000-40 & ISO 7536 | Petroleum products. Determination of oxidation stability of gasoline. Induction period method. |
BS 2000-138 & IP 138 | Methods of test for petroleum and its products. Determination of oxidation stability of aviation fuel. Potential residue method |
Details
Capacity | 60 litres |
---|---|
Fluid type | Water |
Temperature | Bath: 100 °C
|
Max no of test vessels | 4 (Not included with the instrument) |
Vessel charging | 1380 kPa (200 psi) maximum |
Vessel cooling | Cold water plunge tank |
Pressure range/Accuracy | 0-1600 kPa (0-16 bar)/0.1% of span |
Max Test duration | 120 hrs / 7200 mins |
Calibration | Pressure and Temperature sensing with traceable certificate |
Safety & protection | CE certified / IP 67 rated |
Software operating requirements | Windows XP,Vista,7, or 8 (32 and 64 bit) USB or LAN Printer (optional) |
Power | 3 kW |
Voltage | 220/240 V, 50/60 Hz |
Size (HxWxD)/Weight | 93 cm x 86.5 cm x 71.5 cm / 118 kg |
CCCN Code | Tariff 90268020 |
Methods
Spares and accessories for Seta Auto-Oxidation Stability Bath - 15400-5
Water Purifier - 15650-0
View accessoryAnalytical Balance - 99100-2
View productThermometer similar to ASTM61C: Low Haz with Works Certificate - ASTMST61C
View accessoryLaboratory Drying Oven - 99200-3
View productTemperature: ASTM D525 Calibration - 15804-0
View accessoryThermometer ASTMS22C: Low Haz with Certificate of Conformity - ASTMS22CX
View accessoryWant to know more?