Product
AvCount Lube Particle Counter - SA1900-0
The AvCount Lube particle counter determines particle concentration in oils with a viscosity up to 200 mm2/s. FFKM seals have been fitted to resist aggressive oils.
• Simple operation
• Ideal for handling lubricating and hydraulic oils
• 3 minute test time
• Programmable via PC
With a simple test procedure and user friendly features, operator skill or expertise is kept to a minimum.
AvCount Lube can hold up to 3 test methods with custom methods created on a PC before being uploaded using the supplied ProTrend control and analysis software.
Test progress and results for up to 6 size bands are displayed as cumulative particles/ml and ISO 4406, NAS 1638, AS4059 cleanliness codes.
Supplied with: sample probe, waste pipe, universal voltage power supplies, USB cable, calibration certificate and USB memory stick containing handbooks and ProTrend software for test method programming and data download.
AvCount Lube Particle Counter - SA1900-0
Features
- Calibrated to ISO 11171, traceable to NIST SRM 2806d
- Supplied with ASTM D7647, ISO 60970 and ISO 4406 loaded as standard
- FFKM seals compatible with synthetic oils
- Suitable for fluids up to 200mm2/s viscosity
- Precise and reliable results
- Ultra compact rugged, steel case suitable for portable use
- Data storage for up to 600 measurements
- Internal Double Pump System
- USB connection for PC connectivity
Information
Details
Integral CompressorAvCount Lube comprises a light extinction automatic particle counter and a sample delivery unit. Designed to handle samples with a viscosity up to 200 mm2/s, the AvCount Lube has an integral compressor that pressurises the sample chamber to 3 bar.
Calibration and verification:
AvCount Lite utilises 16 point calibration. Calibration can be carried out either using NIST traceable standard solutions or by interfacing with a primary calibrated ‘Master’ AvCount Lite. Calibration can be carried out at a user site or at our authorised regional laboratories. Seta authorised training is required to perform a calibration. Verification can be performed using AvCount Verification Material.
ProTrend (SA1810-0) is supplied with the instrument and can report results to ISO 4406, NAS 1638 and SAE AS4059 (tables 1 and 2) cleanliness codes. Using the software, you can program other test methods and cleanliness codes with up to 16 size bands.
Details
IP 565 | Determination of the level of cleanliness of aviation turbine fuels – portable automatic particle counter method |
---|---|
ASTM D7619 | Method for counting and sizing particles in light and middle distillate fuels, and liquid bio fuels, by automatic particle counter |
GJB 420-1987/420-A-1996/420B-2006 | Fluids contamination testing |
ASTM D7647 | Standard test method for automatic particle counting of lubricating and hydraulic fluids using dilution techniques to eliminate the contribution of water and interfering soft particles by lght extinction |
ISO 60970 | Insulating liquids - Methods for counting and sizing particles |
ISO 11500 | Hydraulic fluid power - Determination of the particulate contamination level of a liquid sample by automatic particle counting using the light-extinction principle |
ISO 4406 | Hydraulic fluid power - Fluids - Method for coding the level of contamination by solid particles |
IEC 60970 | Insulating Liquids, Methods for Counting and Sizing Particles |
JIS B 9932:2012 | Calibration method of automatic particle counter for hydraulic pressure-liquid |
JIS B 9933:2021 | Hydraulic-solid particle contamination degree code display |
JIS B 9934:2012 | By automatic particle counting method using hydraulic pressure-light shielding principle Cleanliness measurement method |
ASTM D6876 | Standard Test Method for Particle Count in Mineral Insulating Oil using Automatic Optical Particle Counters |
Details
Particle Size Range | ISO 11171: 4µm(b) to >70µm(b) |
---|---|
Number of Measuring Channels | 15 when connected to a PC, 6 when used as a stand-alone |
Counts Per Measurement (Max) | 50000/ml |
Sample Viscosity (Max) | 200 mm2/s |
Sample Temperature Range | 0 to 70 °C |
Results | 600 internal, unlimited on PC |
Pump Volume & Flow Rate | 10 ml aliquots @ 30 ml/min or timed |
Total Sample Volume Used (typ) | 80 ml (includes rinse cycles) for ASTM D7647, 20 ml other methods |
Voltage | Automatic particle counter: 12 Vdc, 2.5A (mains adaptor supplied 100 to 240 Vac, 50/60 Hz, 650mA) Pressure delivery unit: 24 Vdc, 2A (mains adaptor supplied 100 to 240 Vac, 50/60 Hz, 1A) |
Size (HxWxD) | 50 x 32 x 28 cm |
Weight | 16 kg |
CCCN Code | Tariff 90318020 |
Methods
Industries
Spares and accessories for AvCount Lube Particle Counter - SA1900-0
Download product profile
To view our full range of products, download our latest Product Profile here.
Want to know more?
View our full range of products, or simply get in contact with us for more information.